![]()
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Energy settings for mono- and biphasic defibrillation:
guideline of the New Zealand Resuscitation Council
Duncan Galletly, Peter Larsen, Nigel Lever, Richard Aickin,
and Warren Smith
The New Zealand Resuscitation Council (NZRC) has received
many requests from health professionals in New Zealand to provide a clear
guideline regarding the energy settings to be used for the defibrillation of
ventricular fibrillation using biphasic
devices.
On the basis of available evidence, the following rationale
and recommendation were developed by the defibrillation advisory group of the
NZRC, and were approved by the NZRC in September 2003.
IntroductionFor a person suffering a ventricular
fibrillation arrest, the key determinant of survival is the time interval
between collapse and the delivery of a defibrillating
shock.1 International guidelines, emergency
medical services, and (now) public access defibrillation programs are all based
upon this evidence. To achieve successful defibrillation, sufficient electrical
current needs to flow through the myocardium,2
but this need must be balanced by the caution that excessive current may cause
direct injury to the myocardium.3
The shock output from defibrillators is measured as Joules
of energy. Currently, the internationally agreed recommendation for
monophasic waveform shock energies (in
the treatment of ventricular fibrillation and unstable ventricular tachycardia)
is an initial escalating triplet of shocks from 200, 200 (or 300), then 360
Joules, followed by further shock triplets maintained at 360
Joules.4 Although long experience has shown
that this energy sequence is clinically effective1,4
there is no conclusive evidence to indicate that this sequence is indeed
the optimal for maximising the number of victims who survive VF/VT
arrest.5,6 The uncertainty in the optimal
energy sequence is due primarily to a paucity of systematic research, which in
turn is due (in part) to the significant ethical difficulties of resuscitation
research.
In recent years, a range of defibrillators delivering
biphasic shock waveforms have become
available. Compared to monophasic
waveform shocks, research clearly indicates that
biphasic shocks achieve defibrillation
rates equivalent to those of monophasic shocks, but at lower
energies.6,7 A theoretical advantage of
low-energy defibrillation is that it may result in less myocardial injury caused
by the direct effects of the electrical
current,8 but, in the clinically used range,
the relevance of myocardial injury is not conclusively
known.4
Biphasic waveforms
vary in their shape according to the electronic method used in their production.
The shape of the biphasic waveform used
in a particular defibrillator differs according to manufacturer, and it is
probable that the energy required for each
biphasic waveform shape to be optimally
effective differs according to that
shape.9,10
Just as with
monophasic waveforms, the ideal
sequence of energy settings for biphasic
defibrillators (used for a population of VF / unstable VT victims) is not
presently known.4,6 Unfortunately, there are no
dose-response studies to define optimal energy sequences based upon large-scale
systematic study of human victims of ischaemia-induced ventricular fibrillation,
and it is on the basis of less conclusive evidence that different energy
sequences have been recommended by manufacturers for different devices
delivering different biphasic
waveforms.4,6
Although there are competing claims for particular
biphasic waveforms and energy
sequences, to date there is no conclusive evidence that any particular brand of
biphasic device or waveform (or energy
sequence), results in greater hospital-discharge rates of neurologically intact
victims of human VF / unstable VT, than any other
brand.4,6,11
Unfortunately, wide variations in manufacturer
recommendations have the potential to create considerable confusion when a
rescuer is faced with an unfamiliar defibrillator (of unknown waveform) during a
cardiac emergency—and may create educational problems in attempting to
provide a simply remembered guideline for recommended energy sequences.
Furthermore, although low (120–200 Joule) energy
biphasic shocks appear to be more
effective than high energy (200–360)
monophasic
shocks,4,6 it is not conclusively known
whether high energy biphasic shocks are
more effective than low energy biphasic
shocks—and it is therefore impossible to state whether
biphasic settings should optimally be
lower than (or indeed similar to) the present
monophasic
settings.6
RecommendationWith the above considerations in
mind, and because of the absence of conclusive human research evidence, the
following recommendation is made by the NZRC. (We emphasise that the primary aim
of this recommendation is to promote safe practice while further research
evidence becomes available; it does not imply that higher shock energies are
proven to be more effective).
The NZRC recommends that, for
manual defibrillation, mono- and
biphasic energy settings (for the
treatment of VF / unstable VT in the adult), the initial settings should be 200,
200 Joules, followed by a third shock at the maximum available energy setting
(up to a maximum of 360 Joules). All subsequent shocks should be given at the
maximum available energy setting (up to a maximum of 360 Joules). If the rescuer
is using a manual defibrillator that is not capable of delivering energies of
200 Joules, it is recommended that the maximum energy output be used throughout
the resuscitation procedure.
In infants and children, it is recommended that
monophasic and
biphasic energy levels (using manually
operated defibrillators) be given with initial escalating settings of 2, 2, and
4 Joules/Kg, followed by subsequent shocks at 4 Joules/Kg.
This NZRC recommendation provides:
The
implications of this recommendation, for the major brands of defibrillator
available in New Zealand, are shown in Table 1.
Table 1. Energy settings for three major brands of
biphasic defibrillator available within
New Zealand, used in accordance with the NZRC recommendation
Author
information: Duncan Galletly; Peter Larsen; Nigel Lever; Richard Aickin;
Warren Smith, NZRC Defibrillation Advisory Group, New Zealand Resuscitation
Council, Wellington
Correspondence:
Peter Larsen, NZRC Defibrillation Advisory Group, New Zealand Resuscitation
Council, PO Box 7343, Wellington. Fax: (04) 389 5318; email: peter.larsen@wnmeds.ac.nz
References:
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Current
issue | Search journal |
Archived issues | Classifieds
| Hotline (free ads) Subscribe | Contribute | Advertise | Contact Us | Copyright | Other Journals |